Manual
– 9 –
6.0 STORING AND RECALLING TENSION VALUES
The tension meter features a data logger which stores the followng data obtained during 
a measuring period:
Average value 
Last value 
Maximum value (MAX), 
Minimum value (MIN), 
Minimum peak value (MIN PEAK) 
Maximum peak value (MAX PEAK)
NOTE:  The measured data remain stored in the ETMB memory even after the   
  instrument is switched off.
6.1  Storing Tension Values
NOTE:  The stored tension values remain stored in the ETMB memory even after the 
  instrument is switched off.
1.  Turn off the tension meter. 
2.  Insert the process material. 
3.  Press the MEM key to start the measuring period. 
While the tension data are stored, the MEM indicator blinks on the 
DISPLAY and the currently measured value is displayed.
When you want to end the measuring period, press the MEM key once again. Data 
logging is stopped. The DISPLAY shows MEM and the current reading.
6.2  Recalling Stored Tension Values
1. Be sure that the tension meter is ON.
  NOTE: You can end recall at anytime by pressing the POWER key.
2. Press the RECALL key. The display blinks, showing 
the average value of the measuring period and the 
indicated symbol.
3. Press the RECALL key. The display blinks, showing the 
last value of the measuring period (no symbol).
4. Press the RECALL key. The display blinks, showing 
the maximum value of the measuring period and the 
indicated symbol.
5. Press the RECALL key. The display blinks, showing 
the minimum value of the measuring period and the 
indicated symbol.
3
D  cN
MEM
22
D  cN
23
D  cN
24
D  cN
21
D  cN
– 12 –
7.0 STATIC VERIFICATION OF MEASURING ACCURACY
All tension meters are calibrated with standard materials—such as polyamide 
monolament (PA)—according to the factory procedure. The diameters are given in 
section 1.0. Any difference in process material size and rigidity from the standard 
material may cause a deviation of the accuracy.
In 95% of all industrial applications, the factory calibration has been proven to provide 
the best results and is used for comparative purposes.
NOTE:  Before beginning, make sure that the tension meter is factory calibrated   
  for a vertical material path. The required reference weight is available.
Verication procedure:
1.  Switch on the tension 
meter.
2.  Hang a weight 
which corresponds 
to the tension to be 
measured from the 
measured material, 
vertically, as shown.
NOTE: Pay attention to the correct unit of measure cN or g (gram).
3. Press the LEVER down all the way down and thread the process material through 
the MEASURING and GUIDE ROLLERS as described in section 4.1, then   
slowly release pressure on the LEVER until the GUIDE ROLLERS return to   
their original position.
4. Before verifying the calibration, move the instrument slowly up and down to 
compensate for possible mechanical friction losses and thus ensure repeatability of 
the measurements.
5. The tension value shown on the Display should be equal to the value of the   
suspended weight (pay attention to the measuring units).
If this procedure shows a deviation, you can recalibrate the instrument following the 
directions in Sec 8.0
MEASURING- and 
GUIDE ROLLERS
Process 
material










