User Manual

4
Table 1
Figure 3
(HR12-10RC-10SDL), HIROSE
RS-232C Port
An important feature of the FGV-X family of gauges is
the RS232C communications port. Data and commands
can be linked to a PC or any other device responding to
RS232C signals for storage or further analysis. Table 1
(below) describes the function of each pin of the
connector (HR12-10RC-10SDL) shown in Fig. 3.
PIN#PIN#
PIN#PIN#
PIN# DESCRIPTION
11
11
1 Analog signal output
22
22
2 Analog GND
33
33
3 Receive data (input)
44
44
4 Digital GND
55
55
5 Communication enable
66
66
6 Transmit data (output)
77
77
7 No connection
88
88
8 Compression overload output
99
99
9 Tension overload output
1010
1010
10 Frame GND
RS232C Output SpecificationsRS232C Output Specifications
RS232C Output SpecificationsRS232C Output Specifications
RS232C Output Specifications
Baud rateBaud rate
Baud rateBaud rate
Baud rate 2400, 4800, 9600, 19200 bps (selectable)
Data lengthData length
Data lengthData length
Data length 8 bits
PP
PP
P
arityarity
arityarity
arity None
Stop bitStop bit
Stop bitStop bit
Stop bit One
Commands/Responses
The FGV-X series of force gauges has the ability to
recognize and respond to various commands from
outside peripherals. If for instance the gauge receives a
legitimate command from a PC, it will respond by sending
back the same code indicating to the PC that the
command was recognized. If for example the command
is a request for average data output, it will send the
data and the code NA etc. If the gauge recognizes an
error while it is receiving a command, an error symbol is
transmitted back to the PC indicating the exact nature
of the error.
Commands to FGV-X Gauge
AAcr Tare
ABcr Stop output
ACcr Change to peak mode
ADcr Change to average mode
AEcr Peak reset
AFcr Change units to Kg (g)
AGcr Change units to N
AHcr Change units to lb (oz)
BAcr Data output request (single reading)
BBcr Continuous data output request (10 times/sec)
BB1cr Continuous data output request (20 times/sec)
BB2cr Continuous data output request (50 times/sec)
BB3cr Continuous data output request (100 times sec)
BCcr Model name confirmation request
BDcr Units confirmation request
BEcr Peak data output request
BFcr Minus Peak data output request
cr (cr: carrige return)
Response from FGV-X gauge
NA""""""cr Average data output
NB""""""cr Peak data output
4-digit number with decimal point
sign (+ or –)
NE""cr Model number
02 = FGV-0.5X
03 = FGV-1X
04 = FGV-2X
05 = FGV-5X
06 = FGV-10X
07 = FGV-20X
08 = FGV-50X
09 = FGV-100X
NH"cr Unit
0 = N
1 = Kg (g)
2 = lb (oz)
Error Symbols
OBcr Command error
OEcr Parity error
OFcr Format error
OGcr Summing error
OHcr Overrun error