User guide

TI-45N Operating Instruction Guide
11
7. Press MODE key. Returns to this Measurement
mode.
6.1 Acoustic Velocity Modes
C = Preset
A = Adjust to any velocity
6.2 How to change Acoustic Velocity
1. Press MODE key 4 times until C followed by 5930 is shown.
2 Press and keys to toggle to preset velocities.
3. Press MODE key 5 times until A then 5930 is shown. This mode will allow
any Acoustic velocity to be set. See section 7.3 for procedure for setting
gauge to a material that is not a preset material or is not listed in the
Acoustic Velocity Chart.
inch
F
5930
6260
10000
4700
1000
1900
2700
4170
4700
5570
5790
5930
5790
5570
4700
4170
2700
1900
1000
4707
10000
6260
Down
Up
TI-45N Operating Instruction Guide
14
6.5 Setup Of Gauge When Acoustic Velocity Is Unknown
In applications where the type of material is not known or the material type
is not listed in the Acoustic Velocity Selection Table the following steps can
be performed to set the gauge to measure accurately.
1. Measure a sample of the material using a micrometer, caliper or similar
measuring device.
2. Turn the power on by pressing the POW key.
3. Place a small amount of couplant fluid on the sample and place the probe
on the sample.
4. After a thickness value is shown on the LCD display along with the Good
Coupling Indicator, remove the probe from the sample.
5. Use the
and keys to increase or decrease the displayed value until
it equals the known thickness of the sample as measured in Step 1. When
the
key is pressed the UP indicator will be show on the LCD display.
The LOW indicator will be shown for the key.
6. The acoustic velocity setting will be automatically adjusted to the correct
value for this material after Step 5 is completed.
7. Change modes to either Mode C or Mode A to display the current acoustic
velocity, as described in section 6.4. Write this value down so it can be
re-entered without having to repeat the above procedures should you need
to change the acoustic velocity to another setting.
6.6 Changing Acoustic Velocity Settings
After determining the required acoustic velocity for the material to be measured,
the acoustic velocity must be adjusted by:
Selecting one of the ten (10) preset values [Mode C]
Adjusting the velocity to the desired setting [Mode A, Fine Adjustment]
METHODS FOR CHANGING ACOUSTIC VELOCITY SETTINGS
MODE Select one from Use or Last Setting 1900 2700
C a list of 10 keys to index 4170 4700 5570 5790
Preset Values thru selections 5930 6260 7390 10000
MODE Fine Adjustment Use or Any acoustic value can be set
A to desired value keys to increase from 1000 to 12000 m/sec
or decrease value