User Manual

5 Operation of the NucleoCounter
24
5.5.4 Instrument Info (F100)
When typing “F + 1 + 0 + 0 + ” the Instrument Info will appear in the display as in the
Display Text Type shown below.
ChemoMetec A/S
NC SCC-100
v4.xx 2005-MM-DD
S/N: 002-01
Figure 16. Instrument Info
The display shows the instrument type in line 2 of the display. The version number and
date (year-month-date) of the internal software of the instrument is shown in line 3.
The serial number of the instrument is shown in line 4 of the display. The display text is
almost identical to the start-up display text (
Figure 8
).
5.5.5 Saving parameters
The settings changed through the user function are in effect immediately. Some of the
settings, e.g. contrast are not automatically saved as the future settings of the
instrument, but in that case the user is prompted for confirmation upon shutting down
the instrument. Status such as date, time and counter are saved automatically. Thus
when the NucleoCounter is turned off, after changing status not saved automatically,
the display will read “Save parameters?”, "ENTER: Save" and "ESC: Cancel".
Save parameters?
: Save
ESC : Cancel
Figure 17
The shut-down display when any parameters has been changed.
By pressing the Enter key <
> the status will be saved and will be in effect next time
the NucleoCounter is switched on. Pressing Escape <Esc> will not save the status and
the instrument will use the previous settings next time it is switched on.
5.6 Zero Count Check (F50)
The instrument can perform a so called
Zero Count Check
(only firmware v 3.04 or
higher). This check is carried out without cassette in the insertion slit of the instrument