5.5. Values Window
The values window is displayed after the initialisation of the LDplus
control system. It is divided into two areas, the left-hand values
window and the right-hand values window.
The left-hand and right-hand value windows are structured as follows:
The up key
is used to select the measuring channel to be displayed
in the left-hand values window. The down key
is used to select the
measuring channel in the right-hand values window.
The following measuring channels are available:
Measurin
g
channel
Set value
Unit of the measured value
Actual value
Total time
Counts the time of the entire process run
Section time
Counts the time of an individual phase (freezing, warm-up
vacuum pump, etc.)
Ice condenser temperature
Temperature of the ice condenser
Vacuum in mbar
Vacuum in the drying chamber. This measuring channel
can only be selected if a vacuum sensor is installed.
Vacuum converted into °C
Vacuum in the drying chamber in °C, converted according
to the vapour pressure curve above ice for water (see the
chapter "General Information on Freeze Drying"). This
measuring channel can only be selected if a vacuum
sensor is installed.