Specifications

5.6.3. Timer
If the timer is active for the phase (timer set value >= 1 minute), the
system inquires before the selected time is over as to whether the
user wants to continue with the next phase.
If the last phase is active, the system inquires as to whether the
process should be continued or stopped (standby).
To deactivate the timer, select an infinite set value (
).
As an option, it is possible to change over to the next phase
automatically when the preselected time is over.
This applies only to a phase change from:
Freezing  Warm-up vacuum pump
Main drying  Final drying
The option can be activated under Menu -> Options ->
Settings -> Automatic phase change.
5.7. Main Menu
The main menu can be activated by pressing the right-hand softkey
Menu
in the active values window.
The following section explains all menu items in detail:
Focus
Scrollbar
Menu title
Menu list
Softkey for
quitting the
menu
Softkey for
opening a
menu item
5.7.1. Change set values
Used to define the phases freezing, main drying and final drying.
First select the phase.
Operating Manual Freeze Dryer ALPHA 1-2 LDplus LDplus Control System 41