Pressure difference = actual vacuum value at the beginning of the
measurement - current actual vacuum value
Fig.28 Intermediate valve between the drying chamber and the ice condenser
Press to finish the pressure increase measurement. The
values for the pressure difference and the measuring time are
frozen. They are indicators for the current amount of water vapor
and thus for the residual moisture content of the product.
The pressure increase measurement stops automatically when the
test time is over or when the maximum permissible pressure
increase is exceeded.
Open
intermediate valve
Close
intermediate valve
5.4.6. Preselection
Preselection for pressure control and air injection.
All preselections will be reactivated after:
a phase change
a program has been loaded
a switch-over from “Run” to “Standby”
Fig. 29 Preselection menu
54 LSC Control System Operating Manual Freeze Dryer ALPHA 1-4 LSC / ALPHA 2-4 LSC