Code 04
Excess temperature “ice condenser”. Temperature
“ice condenser” is 65°C or more! Freeze Dryer was
switched into Standby.
This error occurs when the temperature of the ice condenser
exceeds +65°C. At the same time, all active units are switched off,
i.e. the system is automatically switched to standby mode. The
error message may occur when the temperature of the ice
condenser exceeds +65°C when defrosting the system. The
refrigeration unit is switched off in order to avoid an operation or a
start at too high temperatures. The unit cannot be reactivated until
the ice condenser temperature is lower than +65°C and after
confirmation of the error message.
Code 21
See chapter 5.6.1.
Code 22
See chapter 5.6.1.
Code 23
AC power failure
A power failure is detected if the voltage supply is interrupted
under operating conditions (running mode). The process is
resumed when the voltage supply is available again.
Code 24
See chapter 5.6.1.
Code 25
See chapter 5.6.1.
Code 26
See chapter 5.6.1.
Code 27
Note: Invalid Control System
Please contact our service department!
Code 28
Note: Invalid System Module
Please contact our service department!
Operating Manual Freeze Dryer ALPHA 1-4 LSC / ALPHA 2-4 LSC Troubleshooting • 77