User's Manual

alternate of thermocouple by pressing K/J key switch.
5.
K/J Main display Input Thermocouple Type Select
The K/J key switches the T1 input to select the
K-type or J-type thermocouple as input, when main dis-
play input selected T1. When the thermometer is turned
on, it is set to the type selected that was in use when the
thermometer was last turned off.
6. MIN MAX with Time record mode
(only Main display)
Press MIN MAX key to enter the MIN MAX re-
cording mode, (displays the Maximum reading with
time, Minimum reading with time and Average reading
stored in recording mode). In the this mode the auto-
matic power-off feature is disabled and
key, °C/°F
key, REL key, SET key, Hi/Lo Limits key and main
display T1 T2 T1-T2 key, K/J key are disabled. The
beeper emits a tone when a new minimum or maximum
value is recorded.
Push MIN MAX key to cycle through the MAX,
MIN and AVG readings. If an overload is recorded, the
averaging function is stopped. In this mode, press
HOLD key to stop the recording of readings, all values
are frozen, press again to restart recording.
To prevent accidental loss of MIN, MAX and AVG
data, in this mode can only be cancelled by pressing and
hold down the MIN MAX key for 2 seconds to exit and
erased recorded readings.
7. REL Relative mode (only Main display)
Press REL key to enter the Relative mode, zero the
display, and store the displayed Reading as a reference
value and annunciator REL is displayed. Press REL key
again to exit the relative mode. The relative value can
also be entered by the user. (See “SET mode” later in
this manual). When the desired Relative value has been
entered, press REL key to enter the Relative mode,
press SET key use set Relative value as a reference
value. Press REL key again to exit the relative mode. In
the Relative mode, the value (can not >±3000.0 counts)
shown on the LCD is always the difference between the
stored reference and the present reading.
8. Sec. Minu. Selecting the Time scale
Read the third displayed (Time) in either second or
minute. When the thermometer turned on, it is set to
second. To change the Time scale, press sec. Minu. key.
Maximum Time reading is 100 hours. If 100 hours is
exceeded, reset Time to zero and restart.
9.
SET mode (Relative value set, Time set and
Hi/Lo Limits value set)
9.1 Press SET key to enter Relative values SET mode
(Press ENTER key can escape relative values set
mode), REL set mode. = = = =.= is displayed in
main display. Relative value is entered via overlay
numbers, then press overlay ENTER key, stored the
relative value, enter Time set mode.
9.2 Time set mode, (Press ENTER key can escape Time
set mode) =.= = = : = = is displayed in second and
third display. Time (hours, minutes, seconds) value
is entered via overlay numbers, then press overlay
ENTER key. Time starts from set time value, enter
Hi/Lo Limits value set mode.
9.3 Hi Limit value set mode,
is displayed (Press
ENTER key can escape Hi Limit value set mode), =
= = =.= is displayed in main display, Hi Limit value
is entered via overlay numbers, then press overlay
ENTER key, stored the Hi Limit value, enter Lo
Limit value set mode (Press ENTER key can escape
Lo Limit value set mode). = = = =.= is displayed in
main display, Lo Limit value is entered via overlay
numbers, then press overlay ENTER key, stored the
Lo Limit value and exit SET mode.
9.4 When the thermometer is turned on. The Relative
set value and Hi/Lo Limits set value that was in use
when thermometer was last turned off set values.
10. T1/T2 T1-T2 second display Input Selection
The input selection indicates which input is selected
for second display; T1 thermocouple, T2 thermocouple
or the difference between the two thermocouples
(T1-T2), when the thermometer is turned on, it is set to
T2 input can select alternate of thermocouple by second
display K/J key switch.
11.
K/J Second display Input thermocouple type select
The K/J key switches the T2 input to select the
K-type or J-type thermocouple as input, when second
display input selected T2. When the thermometer is
turned on, it is set to the type selected that was in use
when the thermometer was last turned off.
12. Hi/Lo Limits mode (only Main display)
Press Hi/Lo Limits key to enter the Hi/Lo Limits
comparative mode,
is displayed. When input tem-
perature value exceed Hi or Lo Limits value. The beeper
emits a continuity pulse tone, press Hi/Lo Limits key
again to exit the Hi/Lo Limits mode.
13. WIRELESS MODE: (802W/802UW)
Pressing key over two seconds to start wireless
function and pressing
again for another two seconds
to stop wireless function. The wireless mode will shunt
down if there is no wireless signal for two minutes.
SET CH/ID to 00,00, pressing Hi/Lo Limits key and
power key over six seconds when the meter shuts
down sand then the meter will set channel and ID to
00,00 status. The second display will appear 00, which
means that the channel and ID has been set to 00.
OPERATOR MAINTENANCE
WARNING
To avoid possible electrical shock, disconnect the
thermocouple connectors from the thermometer before
removing the cover.
Battery Replacement
1. Power is supplied by 4pcs 1.5V (SIZE AAA) UM-4
R03.
2. The “
” appears on the LCD display when replace-
ment is needed. To replace battery remove screw from
back of meter and lift off the battery cover.
3. Remove the battery from battery contacts and replace.
4. When not use for long time remove battery.
5. Don’t keep in place with high Temp, or high humidity.
Cleaning
Periodically wipe the case with a damp cloth and deter-
gent, do not use abrasives or solvents.
*Software operation manual is in the software disk.
V1. 071607