3.5.- VOLTAGE CABLES
To measure voltage, you must use 600 V CAT III double-insulated connection cables.
The Kit B_MYeBOX 150 and Kit B_MYeBOX 1500 kits come with the necessary cables:
- UL 600 V CAT III double-insulated voltage cables, or higher.
- UL 600 V CAT III crocodile clamps
Coloured markers are included with the devices to identify the measuring channels according
to each country's standard.
Table 4: Cable colours: European (IEC 60445 :2010)�
Phase Cable colour
L1 Brown
L2 Black
L3 Grey
N Light blue
I Leak Maroon
Earth Green / Yellow
3.6.- CURRENT CLAMPS
The current may be measured using current clamps or transformer clamps.
The device automatically recognises the clamps that are connected to it and shows the neces-
sary parameters in the setup menu. (“6.1.- SETUP MENU: MEASURE SETUP)
Using clamps is necessary IEC 61010-2-032.
Phase and neutral current measurement:
Table 5: Clamps and transformers for measuring the phase current and the neutral current�
Type Scale Measurement range Accuracy
(1)
CPG-5 - 0.05 ... 5 A 0.2% (3 % ... 120% In)
CPG-100 - 1 ... 100 A 0.2% (3 % ... 120% In)
CPRG-500 - 1 ... 500 A 0.2% (3 % ... 120% In)
CPRG-1000 - 1 ... 1000 A 0.2% (3 % ... 120% In)
CPG-200/2000
LOW
1 ... 200 A 0.2% (3 % ... 120% In)
HI
10 ... 2000 A 0.2% (3 % ... 120% In)
FLEX-Rxxx
LOW
100 A 1% (10 % ... 200% In)
MEDIUM
1000 A 1% (10 % ... 200% In)
HI
10000 A 1% (10 % ... 200% In)
Transformer ���/ 0�333V
-
1% ... 200% In
1% (1% ... 19% In)
0.5% (20% ... 120% In)