User's Manual
Table Of Contents
- Appendix I: Sample Bill of Materials (BoM)
- Appendix K: RFS System Test (Cable & RFS Sweep)
Navini Networks, Inc. Ripwave Base Station I&C Guide
133
Step 5. The marker value should be equal to the RFS Only Tx insertion loss within ±2.0 dB,
per the manufacturer’s data. If the insertion loss results do not agree with the
manufacturer’s data, check the test setup.
Caution: An RFS with results greater than the ±2.0 dB limits should not be installed, as a
potential hardware fault exists. Contact Navini Networks Technical Support.
Step 6. Record the data in the RFS System Test Form under “RFS TX PATH LOSS (RFS
ONLY)”. Ensure that the information is recorded under the channel number of the RFS
antenna that is being tested.
Step 7. Repeat steps 5 and 6 for the remaining seven antenna inputs on the RFS.
Step 8. Change to the next test frequency (refer back to Test Setup). Perform steps 1 through 8
until the RFS has been successfully tested at the frequencies shown in Table K4.
RFS Only Receive Verification
Step 1. Ensure calibration of the test setup and RFS Test Box setup for RFS Only has been
performed each time the test frequency is changed.
Step 5. Switch the RFS Test Box to the Receive (Rx) mode.
Step 6. Connect the cable from the Signal Generator to the RFS cal connector. Use a barrel
connector to change the gender, if required.
Step 7. Connect the cable from the Spectrum Analyzer to the RFS antenna input number 1.
Use a barrel connector to change the gender, if required. See Figure K9.
Figure K9: RFS Only Rx Verification
• Connect the Signal Generator to
the Cal Port
• Connect the Spectrum Analyzer
to the first Antenna Element Port
• Connect the Power & Data Cable
to the Test Box
Signal Generator
Spectrum Analyzer
Test Box
(RX position)
• Connect the Signal Generator to
the Cal Port
• Connect the Spectrum Analyzer
to the first Antenna Element Port
• Connect the Power & Data Cable
to the Test Box
Signal Generator
Spectrum Analyzer
Test Box
(RX position)










