REF English INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR CH-406D CITIZEN is a registered trademark of Citizen Holdings Co., Japan. CITIZEN es una marca registrada de Citizen Holdings Co., Japón. Design and Specification are subject to change without notice.
GENERAL REMARKS English General remarks on blood pressure and blood pressure measurement 1. To deflate the cuff quickly, press the “START/STOP” switch. 2. The attached cuff is suitable for arm circumferences of 23-33 cm (9-13 inches). 3. When attaching the cuff, the air hose should be placed on the inside of your arm on top of the artery. 4. Self measurement is not therapy! In no event should you change the dosage of your medication prescribed by your physician. 5.
7. DO NOT try to measure your blood pressure immediately after drinking coffee or tea or after smoking. 8. Measure your blood pressure when you are relaxed and still. Keep the center of the cuff at the level of your heart and DO NOT move your arm or talk. 9. DO NOT measure blood pressure when the cuff has been on your arm for a few minutes or more. During this time, your arm will have built up a higher blood pressure and you will not get valid reading. 10.
PRECAUTIONS FOR USE AND MAINTENANCE Precautions for use English 1. If you suffer from heart disease, high blood pressure or other circulatory disease, consult your physician before using the monitor. 2. If the cuff pressure feels abnormal or you experience any other irregularity while using the cuff, reduce the pressure immediately by pressing the “START/STOP” switch and then consult the sales outlet where you purchased the monitor. 3.
15. Be sure to use this unit only for measuring blood pressure. DO NOT use it for any other purpose. 16. DO NOT use this unit on infants. 17. Blood pressure measurements taken using this monitor are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standard for electronic or automated sphygmomanometers. 18.
PRECAUTIONS FOR USE AND MAINTENANCE Nature and frequency of maintenance: English This product is designed for use over an extended period of time; however, it is generally recommended that it be inspected every two years to ensure proper function and performance. Protect the nature environment: Please help to protect natural environment by respecting national and/or local recycling regulations when disposing of the battery and the product at the end of their useful live.
STANDARDS BF Classification: - Internally powered equipment - BF type applied part - IPX0 - Not suitable for use in presense of flammable anaesthetic mixture with air or with oxygen or nitrous oxide - Continuous operation with short-time loading This Blood Pressure Monitor complies with the European regulations and bears the CE mark "CE 0197".
COMPONENT NAMES AND LOCATIONS Component names English MODE Switch Air hose connector SET Switch MEMORY Switch START/STOP Switch a Stand-by LCD display Air hose socket Air hose Cuff - Eng 8 -
DEVICE DESCRIPTION English Key to display icons Date display Time display Inflation Deflation Systolic blood pressure display Battery icon Diastolic blood pressure display Error icon Memory number display Pulse display Measurement icon - Eng 9 -
LOADING THE BATTERIES 1 Open the battery compartment cover. English 1 While pressing on the lock section of the battery compartment cover, 2 pull the cover toward you to open it. Battery compartment cover 2 Insert the batteries into the compartment. Insert the batteries, taking care to ensure that the positive m and negative , terminals are positioned correctly. 3 Close the battery compartment cover.
REMOVING THE BATTERIES Open the battery compartment cover. ( See page 10, “LOADING THE BATTERIES”.) 2 Remove the batteries. 3 Close the battery compartment cover. Replacing the batteries • If the appears on the display, or if nothing appears on the display when the “START/STOP” switch is pressed, fit 4 new AA Alkaline batteries. • Do not use rechargeable batteries. • Do not use alkaline and manganese batteries together. • The monitor is powered by the batteries provided.
USING AC ADAPTER (OPTION) English Using the optional dedicated AC adapter for the CITIZEN blood pressure monitor (CH-406D) allows you to measure your blood pressure without having to worry about the amount of remaining charge in the batteries. AC adapter socket 1. Insert the AC adapter plug into the socket on the blood pressure monitor. 2. Plug the AC adapter into a mains power outlet. AC adapter plug • If you plan to use the AC adapter for an extended period of time, remove the batteries.
Your blood pressure varies according to your posture, the time of day and a range of other factors. Ideally, you should measure your blood pressure in the same posture at the same time every day. ■ Taking measurements while sitting 1 Place your arm on a table or similar surface with your forearm extended. 2 It is important to have the arm cuff level with your heart. 3 Open your hand slightly on the table so that your palm is facing up and your fingers are relaxed.
ATTACHING THE ARM CUFF 1 Unroll the arm cuff. English • Unroll the arm cuff and thread the end through the metal hoop so that the side with the Velcro patch is on the outside. 2 Metal hoop Velcro patch Put the arm cuff around your arm so that the air hose is to the left of your palm. • The arm cuff should be put directly against your skin but can also be used over a lightweight sleeve. • If you are wearing a thick top, remove it.
4 Pull the end of the arm cuff outwards so that the cuff is snug around your arm and then secure the end to the Velcro. • Carefully place the cuff on your arm, taking care to use the correct tightness. The tightness is correct if you can readily slide a finger between the cuff and your arm.
SETTING THE CLOCK 1 English Initial state. (When you load the batteries in the unit for the first time, carry out the reset procedure.) • Date/Time setting when the batteries are first loaded into the unit. Date and time: January 1 12:00 2 Setting the Month, Date and Time: The current time and date should be set in the order written below. • Each time you press the “MODE” switch, the data item to be modified changes in the sequence shown below. The selected item flashes on and off.
1 Wrap the arm cuff tightly around your arm. ( See page 14 “ATTACHING THE ARM CUFF”.) 2 Plug the air hose connector into the main unit. • Plug the air hose connector firmly into the air hose socket on the main unit as illustrated. 3 Sit in the recommended position. ( 4 See page 13 “ TIPS FOR OBTAINING ACCURATE MEASUREMENTS”.) Start measurement. • Press the "START/STOP" switch. The air pressure is increased and then slowly decreased. When a pulse is starts detected, the measurement icon flashing.
RECALLING DATA ■ Recalling data Press the “MEMORY” recall switch. English Press the “MEMORY” recall switch again. This recalls and displays first the memory number and then the last set of data stored. DATE TIME SYSTOLIC mmHg DIASTOLIC mmHg MEMORY PULSE/MIN. Press the “MEMORY” recall switch again. This recalls and displays the second last set of data stored. Each time you press this switch, the unit displays the data prior to the last set of data displayed.
KEY TO DISPLAY ICONS Condition/Cause Measurement icon This icon appears when the unit is in measurement mode and flashes when a pulse is detected. Remain quiet while measurement is in progress. Battery icon This appears when the level of remaining charge in the batteries voltage is very low or when the batteries are loaded incorrectly. Replace all four batteries with new batteries.
ABOUT “BLOOD PRESSURE” What is blood pressure? (Systolic) (Diastolic) Artery Artery Left atrium Left atrium Left ventricle The artery wall expands. Contraction Left ventricle Normal state is restored to the artery wall. Dilation Your blood pressure changes all the time. Your blood pressure differs according to your age, gender and a range of other factors.
Q Why is the blood pressure measured at home different from that measured by the physician or at a hospital? A Your blood pressure may vary due to exercise, the ambient temperature or your mental state. When you have a physician or nurse measure your blood pressure, the pressure value is likely to be 10 to 20 mmHg higher than usual due to anxiety and/or stress. Knowing what affects your blood pressure helps you to better monitor your health.
BEFORE REQUESTING REPAIRS OR TESTING Make sure to check the following before sending your blood pressure monitor away to be tested or repaired. English Problem Checks Response Display stays blank when you press the “START/STOP” switch. Check whether the batteries are dead. Replace all the batteries with new ones. Check whether the batteries are installed the right way round (m and , oriented correctly). Load the batteries in the correct direction. No measurements can be taken.
Model Number: Measurement system: Cuff: Arm circumference range: Pressure Measurement range: Pulse Pressure Accuracy: Pulse Pressure Pulse Date LCD displays: Time Icons Switch: Inflation: Deflation: Rated voltage: Exhaust: Power supply: Battery duration: Automatic Power Off function: Memory capacity: Dimensions: Weight: Temperature Operating conditions: Humidity Temperature Storage conditions: Humidity Electric shock protection: Safety Classification: Accessories CH-406D Oscillometric method Soft cuff 23.
REF English INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR CH-406D CITIZEN is a registered trademark of Citizen Holdings Co., Japan. CITIZEN es una marca registrada de Citizen Holdings Co., Japón. Design and Specification are subject to change without notice.