2.4 Device setup and wiring
The wiring strongly depends on the particular station. Clarity station
packages which use an A/D converter (e.g. Colibrick) include a cable set
composed of signal, starting and digital output for connecting the Clarity
station to the chromatograph. Using directly controlled detectors skips the
necessity of having an A/D converter - in such case continue in the
chapter "Clarity Configuration" on pg 9.
2.4.1 Standard cable for the Clarity station
Fig 3: Standard A/D converter cable
l Signal cables
Labeled DET1 to DET4 (according to the number of channels), the
cables are supplied as standard without connectors with only stripped,
tinned endings - red (+), white (-) and shielding (analogue ground).
l Starting (marker) cables
Labeled IN1 to IN4 (according to the number of channels), ended
with an RCA connector. Each starting cable has one cable with free
leads [red (+), shielding (digital ground)] for direct connection to the
chromatograph or a valve and one cable ended with a button for those
cases where there is not a starting contact available and it is
necessary to perform a manual start.
l Cables of the digital outputs
Relay contacts labeled, " OUT 1R" to " OUT 4R" (according to the
number of channels), are ended with free leads. They are used for
synchronizing autosamplers in the active sequence without an AS
Control module.
l At the converter end, the cable ends on a CANNON SUB D 37-pin
connector.