Getting Started 5 Tour through the Clarity station
5.2.4 Linking the calibration to the method
In case of a large number of chromatograms, linking the calibration to
each file separately would be a time-consuming process. To avoid this,
the calibration may be linked to the resulting chromatograms
automatically.
l
Return to the Instrument window and use the Calculation icon to open
the Method Setup dialog directly on the Calculation tab . Alternatively,
you can use other icons such as the Integration , Measurement or
Acquisition icon or any command from the Method menu and then
move to the Calculation tab. All of these sections (and some others) are
part of the template method; thus they are present within the same dialog
but on different tabs.
Fig 18: Method Setup - Calculation dialog
l Use the Set... button to select the calibration file and link it to the
method. Change the parameter in the Calculation field to ISTD.
l Exit the Method Setup dialog using the OK button. In the Instrument
window, use the File - Save Method command to apply this change to the
template method.
l Any chromatograms measured with this template method in the future will
have the actual calibration linked.
l To apply this updated template method to already measured data, go to
the Instrument window and use the Analysis - Batch command.
l Select the files to be reprocessed in the left part of the dialog; multiple files
can be selected by left-clicking them while holding the Ctrl or Shift key.
- 30 -