Datasheet

Getting Started 6 Connecting Autosamplers
Variant B - started by opening the contacts
In the second scheme there is autosampler wiring that conversely waits
for output contacts to be opened. This requires different connection
(marked by a circle).
The OUTPUT and READY contacts are connected in parallel and the
autosampler will start its operation after both contacts are opened.
To have the contact on the INT9 board ( INT7, U-PAD2 or Colibrick)
closed in the initial state, it is necessary to set the Output Initial item in the
Digital Output Control dialog (accessible from the Clarity main window) to
LOW.
Fig 20: Wiring of the autosampler - variant B
Note: The start output mapping of the Clarity to individual digital outputs of the
A/D converter can be set in the bottom-right corner of the System
Configuration dialog (for more details see the chapter Clarity
Configuration on pg 9). A detailed description can be found in the
Reference Guide.
- 34 -