Datasheet

6 Connecting Autosamplers Clarity
6.2 AS + LC set - Active Sequence
In LC systems, the autosampler typically governs the timings. The
eventual pump gradient or detector programs are set independently.
Any autosampler that is used in the Clarity Active Sequence without an
AS Control module must be synchronized with the Clarity station by
cables. The IN
n
starting cable should be plugged into the synchronization
output (INJECTION) of an autosampler and the OUT
n
R cable plugged into
the synchronization input (READY) of an autosampler.
Note: Further information about Active Sequence is available in the Clarity
User Guide.
Fig 21: Wiring of an autosampler in an LC set
Note: The labels on the input and output contacts may vary depending on the
type of the autosampler.
When using additional devices (Detectors, LC Pumps, etc.) it is
recommended to connect these devices independently to other digital
outputs of the A/D board. Each device will then need a dedicated row in
the Event Table to be started or stopped.
Note: When the detector or pump start inputs are connected in parallel to the
Clarity start input, problems are often encountered due to the instrument
grounding.
- 35 -