Datasheet
Getting Started 7 Troubleshooting
▌ c) The detector is not allocated to the Clarity Instrument:
Fig 30: Detector not allocated to the Instrument
Description:
Open the System Configuration dialog from the Clarity window using the System -
Configuration... command and check the tab of the corresponding Instrument -
Instrument X. If it has no allocated detectors, add them.
Solution: In the left-hand list Setup Control Modules select the correct detector connected to
the A/D card you are using and drag it to the corresponding instrument on the
right.
If your A/D card is not in the left-hand list Setup Control Modules, add it using the
Add button and repeat the previous step.
Note: More information on the System Configuration dialog can be found in the
chapter "System Configuration" in the Reference guide.
▌ d) You have a licence purchased for data collection for a smaller number of
Instruments:
Fig 31: Small number of Instruments purchased
Description:
Open the System Configuration dialog from the Clarity window using the System -
Configuration... command and check the tab of the corresponding Instrument -
Instrument X. If the symbol of the curve in the tab header is blue, this indicates an
Instrument cannot be used for data adquisition.
Solution: Check your serial number S/N for example using the command Help - About...
from the main Clarity window.
- 45 -