7 Troubleshooting Clarity
7.3.2 Data Acquisition - Simulated
▌ The title "Simulated" is displayed on the Data Acquisition window.
Fig 32: Data Acquisition Simulated
Description:
The corresponding Instrument only displays the simulated curve (from the file
CHANNX.DTA file), or in other words has allocated a DEMO detector to itself.
Open the System Configuration dialog from the Clarity window using the System -
Configuration... command and check the tab of the corresponding Instrument -
Instrument X . If it only has the detector signals from the DataApex DEMO
detector, it is necessary to configure it once more:
Solution: Remove the Detector X from the DataApex DEMO group from the list of equipment
allocated to the particular Instrument and draw the correct detector from the A/D
converter group you are using from the list on the left side of the System
Configuration dialog to its right side.
If you only have DataApex DEMO detectors in the left-hand part of the System
Configuration dialog and your A/D converter is missing, open the Available Control
Modules dialog using the Add button and add it to the configuration of the station.
Then repeat the previous step.
If this does not help, it is necessary to check and possibly re- install the A/D
converter drivers. For more details on that consult the troubleshooting section of
the A/D converter in its corresponding manual.
- 46 -