User's Manual

Table Of Contents
Chapter 2: Initial Setup and Operation - Wand Mode Operation 2-16
Step 5 Slowly scan the patient from head to toe moving at a rate of 0.2 meters a second (7 inches/sec),
holding the SmartWand 1 to 3 inches above the patient. Follow the onscreen instructions shown
in figure 2-10.
It is important to do all the scans(1-5) in order to most accurately identify potential
retained sponges.(Figure 2-11)
If the wand detects an item retained in a patient, the red light on the wand flashes, and the Wand Mode
screen displays the type and quantity of the item(s). Search the patient for the retained item(s).
Figure 2-11 Patient Scan Procedure
Step 6 When the patient scan is complete, press the
COUNT
MODE button to return to the
Count
Mode
screen. If a retained item was found, place the item into the SmartBucket.
Step 7 Remove the SmartWand from the sterile field. Remove the sterile cover and discard it according
to the standard protocol.
Step 8 Return the SmartWand to its holder and the cable to the cord wrap.