User's Manual

Table Of Contents
Chapter 4: Troubleshooting - System Warnings 4-6
System Warnings
System Warnings are serious conditions that have been caused by misuse of the SmartSponge System. To
correct a system warning condition, remove the full bag of sponges, place a new liner on the Count Out Bucket,
and power cycle the system.
Figure 4-2 Example Warning Message Screen
System Reset - Bucket Limit Has Been Exceeded - Remove Sponges and Power Cycle the System
Case Overload - More Than 500 Sponges Detected - Remove Sponges and Power Cycle the System
CAUSE:
There are 100 or more sponges in the Count
Out Bucket.
ACTION:
Remove sponges or discard the full bag and replace with
a new liner - Power cycle the system.
CAUSE:
The overall sponge count limit for the surgery
has been exceeded.
ACTION:
Manually count used sponges.