User's Manual

2-7
Chapter 2: Initial Setup and Operation
-
System Operation
Figure 2-6 Displaying Sponge Types
Deactivating the System
After the SmartWand has been used to perform a patient scan, it is now ready to be deactivated and set aside
for the next surgical case. Deactivating the system will turn off the radio frequency reader and place the system
into a standby “Sleep Mode”.
Step 1 Remove the SmartWand from the sterile field and discard the wand cover and any recovered
items according to standard protocol.
Step 2 Press and hold the Panel Button for 3 seconds to deactivate the SmartWand-DTX System.
The Display will power off along with the Green LED on the wand.
Step 3 Flip the Power Entry Switch to OFF to completely power off the system, if desired.
Step 4 Clean the entire SmartWand-DTX System according to the procedure in Chapter 3 if necessary.
Notes
The Green LED on the Box will stay illuminated until the Power Entry Switch is
flipped to the OFF position. As long as the Box LED is Green, the system is in
sleep mode. Press the Panel Button to enter “Ready to Wand” mode.