With a BED system:
Step 1. Find a suitable location for the Patient Box to be installed.
Step 2. Apply the AC Power.
Step 3. Press and hold the Power Button for 2 full seconds. The Green LED will light and begin to
flash.
Step 4. Choose between JP3 and JP4 and plug the Sensor unit into the Patient Box.
For a BED Sensor:
Step 5a. Apply weight to the sensor pad to hear the activation beep.
The system is now activated.
For a FLOOR sensor:
Step 5b. #2130 will give a 5-second period for the nurse to leave the room without triggering the
FLOOR sensor. Afterwards, the unit will sound a beep to indicate its activation.
The system is now activated.
O
O
P
P
E
E
R
R
A
A
T
T
I
I
O
O
N
N
:
:
With a BED system:
When the patient lifts weight off of sensor pad,
Response at Patient Unit, #2130:
After the Delay Time has expired, the #2130 sounds its alarm warning beep and
sends out a signal to the #2131
After 5 seconds, #2130 will notify the Nurse Station via its nurse call jack and
the signal will repeat every 2 seconds.
Response at Bell Box Unit, #2131:
Once the signal is received from the #2130, #2131 fast flashes all LEDs and
cycle an alarm tone.
Reset the system:
If the patient returns to the sensor pad and applies weight before the Delay Time
expired:
#2130 stops sounding and transmit alarm signal to #2131
No Nurse call will be placed.
#2131 stops sounding and flashing and reset automatically after 5 minutes
After the Delay Time expired and the nurse responds:
Press the large button on the #2131 to stop sounding and lower the flashing
rate of the LEDs
#2131 will reset automatically after 5 minutes.
Place the patient back in BED and return the pressure on the sensor pad.
#2130 will then reset itself automatically.
To lift weight off of sensor pad, and not activate the #2130,