User's Manual

Press the “HOLD” button on #2130 for one full second.
#2130 sounds one beep to confirm the “HOLD” command and will hold the system
for the Hold Length specified at programming step.
Green LED stops flashing and Orange LED begins flashing
You may press the Power Switch once to cancel the “HOLD” command at anytime
before the Hold Length is expired, or #2130 will sound another beep when the Hold
Length is up to warn the user.
Orange LED stops flashing and Green LED starts to flash
During the Hold Length period, #2130 ignores all signals received from the sensor
After the Hold Length period expired before the patient leaves the bed, in despite,
#2130 sounds one beep to indicate it is back in Monitoring Mode.
To extend the Hold Length:
Press the “HOLD” button once to extend the Hold Length for a time period as
specified at pre-set.
With a FLOOR system:
When the patient leaves the bed and steps on the sensor pad,
Response at Patient Unit, #2130:
Immediately, the #2130 sounds its alarm warning beep and sends out a signal
to the #2131
After 5 seconds, #2130 will notify the Nurse Station via its nurse call jack and
the signal will repeat every 2 seconds.
Response at Bell Box Unit, #2131:
Once the signal is received from the #2130, #2131 fast flashes all LEDs and
cycle an alarm tone.
Reset the system:
The nurse responds before the Alarm Length expires:
Press the large button on the #2131 to stop sounding and lower the flashing
rate of the LEDs
Place the patient back in BED and #2130 will then reset itself automatically.
Press the large button on the #2131 again to reset the system, or the system
will reset automatically after 5 minutes.
The nurse responds after the Alarm Length expired:
Place the patient back in BED and #2130 will then reset itself automatically.
Press the large button on the #2131 to reset the system before leaving the
room.
If not, when it is at this semi-reset stage, #2131 will ignore all alarm signals
sent from the #2130
If not, #2131 will reset automatically after 5 minutes.
To move the patient around and not activate the #2130,
Press the “HOLD” button on #2130 for one full second.
#2130 sounds one beep to confirm the “HOLD” command and will hold the system
for the Hold Length specified at programming step.