User's Manual
6
3. System Configuration
3.1. Entering Programming
Mode
Two CTC-1052 programming modes are
available: Local Programming Mode and
Remote Programming Mode.
<
<
N
N
O
O
T
T
E
E
>
>
) GPRS/3G module will be powered
down when in Programming Mode.
When AC Power resumes or when
exits from Programming Mode,
GPRS/3G module will be powered on
again.
3.1.1. Local Programming Mode
(Optional)
From Idle mode, follow the steps below to
enter Local Programming mode.
Step 1. Plug in the telephone set into
LOCAL PROGRAMMING input
located on the rear side of CTC-1052.
Step 2. Pick up the handset.
RED LED lights up.
Step 3. Enter default ACCESS CODE, 1111
followed by #
.
Step 4. CTC-1052 will emit 2 short beeps and
the red LED starts to flash, indicating
it is in Programming Mode.
Step 5. Proceed to program system by
referring to the Commands in section
3.2 Programming Your 1052.
<
<
N
N
O
O
T
T
E
E
>
>
) Local programming is prohibited when
AC Power fails.
) The first digit of Access Code must be
entered within 15 seconds, otherwise
CTC-1052 will exit automatically.
) Failure to enter the correct Access
Code within 2 minutes will cause
CTC-1052 to exit the Programming
mode automatically.
) To exit Programming mode, enter 99
followed by #
, or place the handset
on hook, or disconnect the
Programming telephone set.
3.1.2. CTC-835 Programmer
CTC-835 is a powerful programming tool (sold
separetely) that features a built-in keypad and
LCD display to help you to program the
medical alarm panels effeciently and
conveniently. It also features Once-for-All
Uploading, which allows you program on a
computer via the supplied Pilot software and
you can then upload all the settings to the
medical alarm panel simultaneously with a
single mouse click.
For detailed usages, please refer to the
Operation Manual of CTC-835 Programmer.
3.1.3. SMS Remote Programming
CTC-1052 can be remotely programmed via
SMS commands. Please refer to section 3.3
SMS Remote Programming for usage.
3.2. Programming Your 1052
z How to enter Commands?
Please make sure CTC-1052 is in
Programming Mode (RED LED flashes
continuously) before trying any CTC-1052
Command Functions. Please refer to
section System Configuration, Entering
Programming Mode.
z Follow the protocol below to command
CTC-1052:
CC
# DTMF
ACK
Function
Selection
#
DTMF
ACK
1. 2. 3. 4. 5. 6.
1. CC = 2-digit DTMF Command
2. # = Termination of the DTMF Command
3. ACK = DTMF Command
Acknowledgement
On CTC-1052:
Short beep
: Command succeeded
Long beep
: Command failed
On the programming phone set: