Service manual

Maintenance
4-7
Error Log
To clear the error log:
1. Power OFF the monitor for several seconds and then power ON again. Enter the Boot menu
when the 3-2-1 countdown appears on the System Startup screen (refer to
Troubleshooting
on page 5-1 for detailed instructions).
2. Clear errors by then selecting c.
3. The log can be printed as well as displayed and cleared via the Service Functions menu.
Display Adjustments (90385)
Brightness and Contrast
To adjust the brightness and contrast:
1. Power ON the monitor.
2. At the Main menu, touch the MONITOR SETUP key, MONITOR CONFIG key, and then the
BRIGHTNESS CONTRAST key.
3. Touch the brightness up and down arrow keys and the contrast up and down arrows to
adjust the display view.
!
Error logs are intended for use by Spacelabs Medical personnel only and do not
necessarily indicate monitor malfunction.
!
Prior to clearing logged errors, always generate a printout of the CMOS error log
and attach it to the monitor’s PM record.
CAUTION:
Avoid dimming the keys to the point that they are invisible. If this happens,
before exiting the menu or powering OFF the unit, turn the brightness up by
touching the place where the brightness and contrast up arrows should be.