User`s manual
102
“ON.” Now, when you press the “YES/ENTER” to accept, the following screen will 
appear: 
Variable Delay Start Setting 
The time format (12 hr AM/PM or 24 hr Military) selected in the BIOMED SETUP Menu 
will apply in this table; and, if Military was selected, the AM/PM field will not appear here. If 
the entire delay start screen is to be aborted, press the “NO” key at the ACCEPT field. When 
the start time and date are entered and the data is accepted after the “YES/ENTER” key is 
pressed, the infusion is set to start at the programmed time. The display then returns to the 
next field in the Pre-Prescription Menu. 
After the entire prescription program is accepted, if the pump is turned OFF and back ON, 
press “YES” at PROGRAM and “YES” to REPEAT Rx. The cursor will automatically 
scroll through the program. No changes will be allowed in the Pre-Prescription Menu during 
this time, thus maintaining the originally programmed Delay Start time. To change the Delay 
Start time, press the “NO/CHANGE” key when the Run Options Screen is displayed: 
Run Options Screen 
This will allow the clinician to go back to review the program and at this point a change can 
be made to the delay start time. 
Note:  If the pump is not connected to the patient’s access site at this time, the patient or 
caregiver should be taught proper techniques to access the infusion site before the 
scheduled start time. 
When the “”RUN” key is pressed, the Green and Yellow LED lights will flash to indicate that 
the pump will automatically start when the delay start count down period is up. During the 
delay start time, the following screen appears: 
If the KVO rate is set to 0, the pump will beep ten minutes prior to the beginning of the 
Continuous Therapy start time. To stop the audible beep, press the “Silence” key. 