Also, when the Air-In-Line sensor is completely disabled and the infusion goes to an 
“Infusion Complete” or “Empty Bag” status, the Air Sensor setting will automatically change 
to 2.0 ml. 
6  AUDIO (Audio Volume): This feature allows you to adjust the audio volume of the pump’s 
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting). 
When you select a number, the pump will beep at that audio level. Press the “NO” key and 
enter a new setting (1 - 9). Repeat until the desired setting appears on the display, and then 
press “YES/ENTER” to select it and move the cursor to the next field. 
7  PWR CK? (Power Check): This feature allows you to periodically recheck the battery charge. 
It will display the active power source of the pump, whether batteries or external power. If 
the active source is the batteries, a graph will indicate approximately how much charge 
remains in the batteries. To access this information, select “ON” and press “YES/ENTER”. 
The Power Source Screen will display for five seconds and return to the Options Menu. 
8  ACCEPT OPT?: If the changes made in the Options Menu are acceptable, press the 
“YES/ENTER” key. If the changes are not acceptable, use arrow keys to cursor up to any 
field for corrections, or abort all changes made in this menu by pressing the “NO” key and 
return to the previous screen. 
There are two remaining selections to consider from the “Run Options Screen”: NO  to 
REVIEW and RUN to START. 
Reviewing or Changing the Program 
To review or make any changes in the prescription programming, press the “NO/CHANGE” 
key at the Run Options screen, and the prescription screens will reappear (Changes may be 
inhibited by lock setting.) Changes to the prescription can be made before the start of the 
therapy; however, once a therapy has started, no changes can be made to the Pre-Prescription 
Screen. It will appear only for review purposes. Attempts to change the Pre-Prescription Menu 
during a new program review will cause an alert screen to be displayed, notifying the programmer 
that changes will require reprogramming of the prescription. 
Starting the Infusion 
RUN:  When all parameters are correct and the pump has been primed properly and connected 
to the access site, press “RUN” at the Run Options screen to start the therapy. If a delayed start 
is programmed, the “Delay Start Count Down Screen” will appear. If no delay start is 
programmed or the delay has elapsed, the pump will begin infusing and the following screen 
displays the status of the therapy: 
Variable Run Screen 1 
This information toggles with the following: 