Interrupting an Infusion 
The infusion can be paused at any time by pressing the “PAUSE” key. The yellow “Standby” 
LED light blinks, and the following screen displays: 
Pause Menu Screen 
The screen offers the user three options: 
Resuming the Current Container 
1  RESUME: To resume the therapy in progress from where it left off, press the 
“YES/ENTER” key at this field. The pump will return to the Run Options Screen, and you 
can resume the therapy. 
Note: If you wish to hang a new container, move the cursor to REPEAT Rx (see following 
section). 
Run Options Screen 
a.  Press the “RUN” key to resume the therapy. 
b.  Press the “NO” key to go back to the programming screen to make changes in the 
therapy program (this feature may be blocked by a security lock, thus not allowing 
changes to the program unless an access code is given). Also, remember that no changes 
are allowed in the first prescription menu if an infusion is in progress. 
c.  Press the “OPTIONS” or “PRIME” keys if either of these functions is desired at this 
time. 
Hanging a New Container 
2  REPEAT Rx: To repeat the present therapy and add another IV bag or reservoir with the 
same prescription, press the “YES/ENTER” key at this field. If more than one therapy is 
programmed, a “Select Therapy Screen” will list all preprogrammed therapies. Select which 
therapy to repeat by moving the cursor to the therapy of choice and pressing the 
“YES/ENTER” key. 
The pump will then display the program screen with the last programmed values of the 
therapy and the cursor will automatically advance through the prescription menus. You can 