User`s manual
118
If no air-eliminating filter is being used, press NO. The infusion will 
start/resume and the Air In Line sensitivity will be set to 2 ml for the 
remainder of the infusion. 
Note:  If the screen is not acknowledged with a YES or NO key press, the 
Air In Line sensitivity will be set to 2 ml when the screen is cleared 
(i.e. alarm occurs while screen is displayed). 
LCD Display:  ALERT AIR SENSOR HAS BEEN SET TO 2.0 ML 
Alert:  Air in line sensitivity will be set to 2.0ml 
Cause:   NO was pressed at the “AIR SENS OFF. Using In Line Filter?” alert 
screen. The air in line sensitivity will be set to trigger an AIR IN LINE 
alarm if 2.0ml of air is detected. 
Resolution:  If an air eliminating filter is being used, press YES when the “AIR SENS 
OFF. Using In Line Filter?” screen is displayed. This will disable air 
detection as indicated under “AIR SENS OFF. Using In Line Filter?” alert 
explanation. 
LCD Display:  ALERT AMT NOT 0 OR CURRENT DOSE AMT 
Alert:  Unable to change clinician dose while clinician dose in progress 
Cause:   A clinician dose is in progress, and a new clinician dose value is entered 
that is not “0” or the current dose amount. 
Resolution:  If the current clinician dose is to be continued, change the ‘Clin DOSE’ 
value back to the original value and accept the field and menu. 
If the current clinician dose is to be terminated, which will allow a new 
clinician dose to be programmed, enter a value of 0 in the ‘Clin DOSE’ 
field and accept the field and menu. The next time that the Options menu 
is accessed a new ‘Clin DOSE’ value can be entered. 
LCD Display:  ALERT AMT TBI GREATER THAN BAG VOLUME 
Alert:  Value entered is greater than value in Bag Vol field. 
Cause:   Occurs when the amount the entered for the “Amt TBI” or “Vol TBI” 
field is greater than the value in the bag volume field. 
Resolution:  Recheck the data entry. 
LCD Display:  ALERT AMT TBI GREATER THAN BAG VOLUME. YES to 
Continue NO to Return 
Alert:  Insufficient volume 
Cause:   Occurs when bag volume is less than amount to be infused. Can occur 
when pump’s priming feature is used as bag volume is depleted by the 
amount used in priming. 
Resolution:  To modify the Bag Volume or Amt/Vol TBI, press “NO” to return to the 
Run Options screen. The programming screens will then be accessible by 
pressing “NO” to Review on the Run Options screen. 










