User`s manual
121
If it is not necessary to preserve the therapy in progress or any other 
programmed therapy, use the “CLEAR Rx” feature of the Clinician 
Biomed menu. This feature will erase all programmed therapies in the 
pump. The Date and Time can then be changed. 
If the therapy is to be terminated but the prescription preserved (excluding 
Delayed Start), go to the “Pause” menu, select “REPEAT Rx” and press 
“YES” at the “Rx in progress” alert screen. The pump can then be turned 
off and then on to access the Date/Time change option in the Clinician 
Biomed menu. Selecting “NEW PROGRAM” and the name for the 
therapy in progress, instead of “REPEAT Rx”, may also be used but this 
selection will erase the prescription for the therapy. 
LCD Display:  ALERT DATE/TIME INVALID 
Alert:  Unable to enter invalid date or time 
Cause:  Date or time entry is not in a valid format. 
Resolution:  Recheck data entry. An invalid date or time will not be accepted. 
LCD Display:  ALERT DELAY START TIME NOT VALID 
Alert:  Delay Start time not valid 
Cause:   Delay start time entered is less than one minute or greater than 24 hours 
from the current system time. 
Resolution:  Reset Delay Start Time to fall within pump limits. (To obtain the current 
date and time of the pump, turn the pump off and then on and observe 
the time on the Welcome screen). 
LCD Display:  ALERT DOOR OPEN or SET NOT PROPERLY INSTALLED 
Alert:  Door Open or Administration Set is not properly installed 
Cause:   Door Latch is not completely closed or the Administration Set is not 
installed according to installation instructions. 
Resolution:  Recheck the door latch assembly and securely close the door. If the door 
was closed properly, re-install the set according to the installation 
instructions.  
LCD Display: ALERT DOSE AMT IS LESS THAN DOSE INFUSED 
Alert:  Dose amount entered is less than the amount already infused for the dose. 
Cause:   New dose value entered is less than the amount already infused for the 
dose. 
Resolution:  Recheck data entry. After the dose in progress finishes, the dose amount 
may be changed to the lower value. 
If the above is not suitable, re-program the prescription. 
LCD Display: ALERT DOSE CANCELLED, RESUME INFUSION, THEN NEW 
DOSE 
Troubleshooting 