User`s manual
133
LCD Display:  UPLOAD CkSum ERR 
Alarm:  CMS Upload experienced a Check Sum Error 
Cause:   At the completion of an upload of a CMS script the Check Sum signature 
from the CMS system did not match the signature calculated by the pump 
Resolution:   A second attempt to upload the script should be made. If the second 
attempt fails, then a new script should be created for the desired 
prescription. (See CMS User’s Manual). 
Troubleshooting 