User`s manual
150
Adapter eliminator and remove one C-Cell battery.) Observe that the Auxiliary 
Alarm System is activated, with the Red LED and Beeper on constant (not 
intermittently beeping). With the battery still out of the pump, press the ON/OFF 
button to turn off the auxiliary alarm. Replace the battery in the battery 
compartment and turn on the pump. Resume the therapy and confirm that the 
infusion continues where it stopped due to Auxiliary Alarm occurrence. 
4  SET BIOMED CLINICIAN MENU ITEMS: 
a  Turn the pump Off and back On. While pump is performing Self Tests confirm that 
the pump is set to the correct Date and Time. If date and time are not correct, 
program the correct date and time from the Biomed/Clinician menu. 
b  Enter the Biomed/Clinician Menu using the Clinician Code. 
c  Set the Lock Level to OFF in the CHANGE LOCK menu. 
d  Clear the Pt DATA from the pump. 
e  Observe the Pump Maintenance Date, top of the Biomed/Clinician Menu, to insure 
adequate time for the next use by patient. 
f  Sign, date and return the Check List to your record file. Pumps that pass can be 
used for new patients. Pumps that fail are to be returned for service. 