User`s manual
10 10
Disposable Administration Set Illustrations and Features 
CAUTION: USE ONLY CURLIN MEDICAL ADMINISTRATION SETS 
WITH THIS PUMP 
Figure 1.3 Samples of Curlin Medical Administration Sets 
These drawings are 
NOT to scale.
Note: An important feature of the Curlin Medical administration set is the unique 
design of the set-based “Integral Flow-Stop.” When the set is new, the “Integral 
Flow-Stop” has a breakaway tab that keeps the “Integral Flow-Stop” open and allows 
the tubing to be gravity primed. When you are ready to load the administration set 
into the pump, remove this tab, and the “Integral Flow-Stop” will automatically clamp 
the tubing. The Curlin Medical “Integral Flow-Stop” prevents inadvertent flow of 
medication to the patient whenever the door of the pump is opened because the 
“Integral Flow-Stop” automatically reacts to clamp the tubing. The “Integral Flow-
Stop” can, however, be opened after the tab is broken away by intentionally squeezing 
down on the movable spring action section of the “Integral Flow-Stop.” See Figure 1.6 
for an illustration of how to intentionally open the “Integral Flow-Stop.” For a 
complete listing of all available administration sets, see the Curlin Medical Product 
Catalog.
Integral Flow-
Stop with 
breakaway tab
T
ubing Guide 
Bag Spike
Slide Clamp
Male Luer 
Lock 
CURLIN
MEDICAL
Integral Flow-
Stop with 
breakaway tab
T
ubing Guide 
Medication 
Reservoir 
Slide Clamp 
Male Luer 
Lock 