User`s manual
15 15
Figure 1.9 Proper installation of the Curlin Medical Administration set into the pump 
Steps in Installing Administration Set into Curlin Medical Pumps 
1  Fully open the door of the pump. 
2  Close the slide-clamp on the administration set. 
3  Activate the Integral Flow-Stop by twisting and removing the breakaway tab. 
4  Insert the blue tubing guide into the receptacle on right side of pump in the direction of the 
blue arrow, positioning the tubing in front of the door hinge. 
5  Center the tubing in the middle of the pumping fingers, covering the yellow dot. 
6  Holding the yellow Integral Flow-Stop by the handle (see Figure 1.9), insert it on an angle into 
the receptacle on the left side of the pump in the direction of the yellow arrow. CAUTION: 
Do not press down on the top of the Integral Flow-Stop when inserting it into the 
receptacle. This action could inadvertently open the Integral Flow-Stop and allow 
fluid to flow through the tubing.  
7  Open the slide-clamp on the administration set 
8  Push the tubing slightly into the black Air Detector slot. 
9  Fully close the door of the pump and latch it securely. 
10  If the set is not primed, prime the set according to prime section in the relevant therapy 
chapter. 
11  Attach the primed set to the patient’s venous access device per agency protocols. 
12  Dispose of used administration sets in accordance with all applicable regulatory and 
institutional policies and directions. 
Introduction 
Air Detector 
Flow Stop 
Handle 
(Yellow) 
Note: Open the 
slide clamp 
before closing 
the pump door 
Tubing 
Guide 
(
Blue
)
Administration Set
Pumping Fingers
Flow Stop 
Rece
p
tacle 
Tubing Guide 
Rece
p
tacle 
Note: The 
Administration 
Set tubing is 
always placed in 
front of the door 
hinge 
Figure 1.9 shows the 
Curlin Medical pump 
with a cut-away of the 
display screen allowing 
y
ou to visualize the 
receptacles for the 
small 
blue 
“Tubing 
Guide” on the right 
side of the pump and 
the larger 
y
ellow 
“Integral Flow-Stop” 
on the left side of the
p
ump. The arrows, in 
corresponding colors, 
are visible on the 
background surface of 
the pump head and 
indicate the proper 
direction for placing 
these locators into their 
receptacles. Also, note 
that the breakaway tab 
has been removed from 
the “Integral Flow-
Stop”, allowing the 
administration set to be 
placed into the pump 
correctly.










