User`s manual
36
11  CLEAR Pt. DATA: Selecting this feature clears all data programmed and resets the 
pump to its default settings. The history log is NOT cleared. All disabled therapies 
are reenabled. When you select “YES/ENTER” at this field, the following screen displays: 
When you select “YES,” all patient data clears and the display returns to the BIOMED 
SETUP Menu. Selecting “NO” aborts the feature and also returns to the menu. 
12  CHANGE CONTRAST: The contrast on the display screen can be adjusted. To do so, 
press the “YES/ENTER” key at this field, and the following screen appears: 
The level can be adjusted from 0 (lightest screen) to 100 (darkest screen). The pump is set to 
an optimized setting when shipped from the factory. To change the setting, enter any 
number between 0 and 100, and press the “YES/ENTER” key to see the change in the 
screen. If this level of contrast is not satisfactory, move the cursor back to the “LEVEL” 
field and enter another number and press the “YES/ENTER” key. When the contrast is 
satisfactory, move the cursor to the ACCEPT field, and press the “YES/ENTER” key. To 
abort the entire selection and return to the previous contrast setting, press the 
“NO/CHANGE” key at this field and return to the BIOMED SETUP Menu. 
Changing Date and Time 
13  CHANGE DATE AND TIME: When this field is selected, the screen on the left appears: 
Change Date and Time Screen 
This screen is used to set the system Real Time Clock in the pump and will change the date 
and time displayed on the Welcome screen. 
Note: The time must be reset to accommodate Daylight Savings Time changes and also at any 
time the pump is moved into a different time zone. 










