If only one therapy is programmed and REPEAT Rx is selected, the cursor will automatically 
advance through the prescription menus. You can stop the auto-advance function by pressing 
any key (except ON/OFF) and then complete the confirmation of the prescription by 
pressing the “YES” key at the remaining fields. 
Note: REPEAT Rx does not allow any changes to the Pre-Prescription Menu.  If changes 
need to be made in Pre-Prescription selections (units, concentration, etc.) select “New 
Program” and reenter the entire prescription. 
The lock settings affect the use of “REPEAT Rx”. Lock 1 allows REPEAT with changes to 
the prescription screen only. Lock 2 allows REPEAT but does NOT allow the user to make 
changes in the prescription. Lock 3 does not allow REPEAT. (See lock setting table). 
If a therapy has not completed and the REPEAT Rx function is selected, the following screen 
will display: 
If the current infusion is to be discontinued and a new IV bag started, press “YES/ENTER” 
to repeat. 
If the current infusion is to continue, press “NO,” and you will be prompted to resume the 
therapy. 
How to go to “New Program” 
3.  NEW PROGRAM—If the cursor is moved to this field and the ‘“YES/ENTER” key is 
pressed, the display changes to allow you to select which therapy you wish to program: 
Selecting a Therapy 
Select Therapy Screen 
This screen displays only those therapies enabled from the BIOMED SETUP Menu. Move 
the cursor to the appropriate therapy and select it by pressing the “YES/ENTER” key. If 
Basics Programming 