User`s manual
45
6  DONE?: If any changes are to be made, use the up or down arrow keys to locate the field to 
be changed, use the numeric keys to enter the corrected information, and press 
“YES/ENTER” to accept the new data. If all the fields are acceptable, move the cursor to 
the “DONE?” field and press the “YES/ENTER” key to move on to the next screen. 
Before the infusion is started, it is important to consider a few 
remaining issues, such as checking the administration set to be 
sure it was primed or purged of all air, or perhaps the pump may 
need to be put into a security lock setting for additional safety. 
The next screen that appears when the programming is complete offers four choices: 
Run Options Screen 
Note that you do not need to answer yes or no questions or enter data with the numeric keys as 
previously requested. This screen requires only that you press the corresponding key on the 
keypad to begin the requested function. The following explains each selection beginning with 
Prime: 
Prime 
The IV bag and administration set should be inspected and be free of all air and air bubbles 
before they are connected to the patient’s access site. Priming may have been done by gravity 
when the administration set was first connected to the bag (see Chapter 1, “Preparing Medication 
for Infusion”). However, if it was not done then or if there are any remaining air bubbles to be 
removed, the pump’s “PRIME” function will assist in this process. To prime the set by using the 
pump, follow the directions below. 
Note:  The lock setting may restrict the ability to prime using the pump. 
1  Press and release the “PRIME” key at the “Run Options” screen. 
2  The following screen appears: 
Prime Direction Screen 
3  Disconnect the administration set from the patient’s access site using aseptic protocols. 
Priming with the set connected to the patient could result in overdose 
and may cause injury or even death to the patient. 
IMPORTANT 
INFORMATION 
Continuous Therapy 
IMPORTANT 
INFORMATION 










