4  Release any clamps on the administration set. 
5  Press and hold the “PRIME” key. Fluid will be pumped at a rate of 300 ml/hr through the 
administration set as long as the PRIME key is pressed or until 6 ml is pumped. The pump 
will then beep once and require repressing and holding the “PRIME” key again until 6 more 
milliliters is delivered. Continue this process until all air has been purged from the IV bag and 
the administration set. 
Note:  If the bag volume is depleted during priming, a subsequent priming event cannot be 
initiated. 
  While the “PRIME” key is depressed, the following screen appears: 
  The amount used for the prime session is accumulated and displayed on the bottom field of 
this screen. 
6  When priming is complete, release the “PRIME” key. The following screen will display: 
7  To continue priming, press the PRIME key again. If priming is complete, press 
“YES/ENTER” to exit and return to the Run Options screen. The total amount of solution 
used in this priming session displays on the bottom field. 
Note: The amount of fluid used in the priming process will be deducted from the Bag Volume 
Total but will NOT be counted in the amount infused because this amount is not to be 
delivered to the patient. The bag volume should still be equal to or greater than the 
Amount TBI before you can proceed with the infusion. If it is not, an alert will appear 
stating, “AMT TBI GREATER THAN BAG VOL”. You have the option to continue 
by pressing the “YES/ENTER” key or return to the “Run Options” screen by pressing 
the “NO” key. 
Options 
Once the prescription is programmed, there are more customizing features available for each 
therapy. Press the “OPTIONS” key from the Run Options screen to access these options. The 
OPTIONS menu can also be accessed while the pump is running. 
Note. After the infusion has started, any changes made through the OPTIONS menu involving: 
•   “HRLY TOT”, and 
•  “CLR HRLY” 
will return you directly to the “ACCEPT OPT?” field. 