User`s manual
56
3  ADMIN Rt (Administration Route): Allows a selection from three routes of administration: 
the default setting of IV (intravenous), EPI (epidural), and SQ (subcutaneous). If Epidural is 
selected, the Therapy Identifier bar in the Prescription Menu will display EPI instead of PCA.  
Each of the routes has defined settings as illustrated in the following table: 
Administration Routes 
Administration 
Route 
Occlusion Setting 
mmHg (psi) 
Maximum 
Basal Rate
(ml/hr) 
Rate During 
Bolus 
(ml/hr) 
Maximum 
Bolus Dose 
(ml) 
Maximum
Load Dose
(ml) 
Intravenous  900 or 400 (18 or 8)  50  125  50  50 
Epidural  900 or 400 (18 or 8)  25  90  25  25 
Subcutaneous  900 or 400 (18 or 8)  5  60  5  5 
Figure 5.1 PCA Administration Route Default Settings  
4  Load Dose (Loading Dose): Allows the clinician to program in a pre-infusion loading dose if 
it is prescribed. If a loading dose is not ordered, select the default setting of “0” by pressing 
“YES/ENTER”. If it is ordered, enter the dose in this field, and press the “YES/ENTER” 
key to accept and move to the next field. The maximum load dose is determined by the 
administration route (refer to Figure 5.1 above). 
Note:  The “units” will correspond to the choice previously selected. 
The loading dose will be delivered at the fixed bolus delivery rate when the “RUN” key is 
pressed to start the PCA infusion. The amount of the loading dose is included in the 
displayed “amount infused” and is logged into the history file. The bolus lockout time is set 
from the end of the loading dose. Because the loading dose is only used on initial start up of 
an infusion, it will revert to a value of “0” whenever the “REPEAT Rx” function is selected. 
To terminate a loading dose in progress but save the rest of the 
prescription, pause the pump and select “REPEAT Rx.” When the 
“RUN” key is pressed, the prescription will restart without the 
loading dose. 
5  Titrate: Allows the clinician to program titration limits into the prescription settings of the 
pump. If there is no titration prescription, select “OFF,” and press “YES/ENTER” to 
accept, and move the cursor to the next field.  If a titration prescription has been written, 
select “ON”, press the “YES/ENTER” key to accept, and the following screen will appear to 
program the titration parameters: 
PCA Titrate Limits Screen 
IMPORTANT 
INFORMATION 