User`s manual
62
Note:  If a lock setting of 1, 2 or 3 is selected, an access code will be requested. If you wish to 
give a clinician dose at this time, change OFF to ON and press YES key. Enter the 
amount of clinician dose to be delivered. Press YES to accept the clinician dose menu. 
The Clinician Dose will begin when the options screen is accepted and the pump is 
running. The following Run Screen will appear until it is completed: 
Clinician Dose Run Screen 
The graphic bar now depicts the clinician dose in progress and when this dose is complete, 
reverts back to reflecting the container in progress. In CMS-enabled pumps, the graphic 
bar will toggle with Line Pressure (“L Press”) which displays the current patient side 
(downstream) pressure in the administration set in millimeter of mercury (mmHg) 
units. 
The Volume Infused will toggle with Dose Infused if mg or mcg are the selected unit. 
Remaining dose time will toggle with remaining dose volume to infuse. 
When the clinician dose is complete, the pump will beep, return to the 
previous running rate, and resume the PCA therapy. The lockout 
time for bolus dosing will restart at the end of the clinician dose to 
prevent a bolus from being delivered too quickly after a clinician dose. 
  Once the clinician dose is given, the clinician-dose field returns to its default of “0,” and any 
subsequent clinician doses will need to be reentered. 
3  SHFT TOT (Shift Totals): This feature allows the clinician to track the infusion for a selected 
period of time, such as an 8 or 12-hour shift. To access Shift Totals, press the 
“NO/CHANGE” key to change OFF to ON and then press “YES/ENTER”. The 
following screens will appear alternately: 
Shift Totals Screens 
a.  Cleared/ELAPSED: This field displays the date and time the shift total was last cleared.  
It toggles with the amount of time elapsed since that clearing. 
b.  Vol INF/Amt INF (Volume/Amount Infused): This field displays the volume (ml) and 
amount (mg or mcg) infused since the shift total was cleared. 
c.  BOLUS/Clin DOSE (Clinician Dose): This field displays the number of clinician doses 
given since the total was cleared. It toggles with the number of patient activated boluses 
received of those requested. 
IMPORTANT 
INFORMATION 










