enter a new setting (1 – 9). Repeat until the desired setting is reached, and then press 
“YES/ENTER” to select and move the cursor to the next field. 
10  PWR CK? (Power Check): This feature allows you to periodically recheck the battery charge. 
It will display the active power source of the pump, whether internal batteries or external 
power. If the active source is the batteries, a graph will indicate approximately how much 
charge remains in the batteries. To access this information, select “ON” and press 
“YES/ENTER”. The Power Source Screen will display for five seconds and return to the 
Options Menu. 
11  ACCEPT OPT? If the changes made in the Options Menu are acceptable, press the 
“YES/ENTER” key. If the changes are not acceptable, cursor up to any field for corrections 
or abort all changes made in this menu by pressing the “NO” key, and return to the previous 
screen. 
There are two remaining selections to consider from the “Run Options Screen”, NO  to 
REVIEW and RUN to START.  
Reviewing or Changing the Program 
To review or make any changes in the prescription programming, press the “NO/CHANGE” 
key at the Run Options screen, and the prescription screens will reappear (Changes may be 
inhibited by lock setting.) Changes to the prescription can be made before the start of the 
therapy; however, once a therapy has started, no changes can be made to the Pre-Prescription 
Screen. It will appear only for review purposes. Attempts to change the Pre-Prescription Menu 
during a new program review will cause an alert screen to be displayed, notifying the programmer 
that changes will require reprogramming of the prescription. 
Starting the Infusion 
RUN: When the programming is complete and the pump has been primed properly and 
connected to the access site, press the “RUN” key at the Run Option screen to start the therapy.  
The pump will begin infusing, and except when a LOAD DOSE or CLIN DOSE is 
programmed, the following screen will display data on the status of the therapy: 
PCA Run Screen 
This screen will continue to display as long as the infusion is running. Notice the information on 
each field: 
1  The first field displays the “RATE” of the infusion toggled with the programmed Bolus Dose 
Amount. This field also displays “BOLUS xx (mg/ml/mcg)” whenever a bolus dose is being 
delivered. The programmed bolus amount displays and the BOLUS text flashes when the 
pump is delivering a bolus dose. 
2  The second field is a graphic bar representing the percentage of the completed infusion. In 