3  CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the “NO/CHANGE” key 
to change OFF to ON and then press “YES/ENTER”. The following screen will appear: 
Clear Hourly Totals Screen 
a.  To clear the Hourly Totals, press the “YES/ENTER” key. All Hourly Totals will be 
cleared and the cursor will return to the Options menu. 
b.  To maintain the Hourly Totals, press the “NO/CHANGE” key. Hourly Totals will be 
maintained and the cursor will return to the Options menu. 
c.  A lock level of 2 or 3 will require an access code entry before allowing “Hourly Totals” 
reset. 
4  DN Occlu (Down Occlusion): This feature allows the pump to sense an occlusion pressure in 
the line between the pump and the access site (e.g., a kink in the tubing between the pump 
and the patient, a blockage in the access site, or a clamp left on the distal tubing). There are 
two options for this field: the “Low” setting (default setting) triggers the alarm when it detects 
in-line pressures at 400 mmHg ±150 (8 ±3 psi “pounds per square inch”), or the “High” 
setting alarms when it detects in-line pressures at 900 mmHg ±150 (18 ±3 psi). Press 
“NO/CHANGE” to change the selection. When the correct pressure setting appears, press 
the “YES/ENTER” key to accept it and move the cursor to the next field. 
5  AIR SENS (Air In Line Sensor): This feature allows you to select the amount of air the pump 
will detect in the administration set tubing before it goes into an alarm state. The choices are 
0.1 ml, 0.5 ml (the default setting), 2.0 ml, and OFF. Press the “NO/CHANGE” key to 
change the options until the desired setting appears on the display, and then press 
“YES/ENTER” to select and move the cursor to the next line.  
Whenever the Air-In-Line Sensor is set to OFF, Curlin Medical 
recommends the use of an air-eliminating filter. 
When the Air-In-Line Sensor is set to “OFF” and the “RUN” key is pressed at the Run 
Option screen, the alert message below will appear on the screen. It will also appear if the 
Air-In-Line is changed from the OPTIONS menu when an infusion is in progress and the 
AIR SENS field is accepted. 
Air-In-Line Sensor Off Message Screen 
IMPORTANT 
INFORMATION 