Technical data
-23- 
Special Status and relevant Error Messages 
The table below describes those special status which cause the yellow fault LED to be lit 
permanently. For a better diagnosis of a transmitter without display you should either read the error 
messages below from the remote control RC2 or analyse the values from the current output. 
No 
Display 
green 
LED 
yellow 
LED 
Output 
Cause  Action 
01 
"
TEST
"
On  On  0 mA  Memory test when booting the system  finishes automatically after 6 
seconds 
02 
"
Load
" 
Operational 
para.
" 
AdJ
 "
flashing  On  1,6 mA  Booting the system at start or after 
change of gas (see page 8) 
turns to sensor warm-up 
automatically 
03 
Countdown 
of seconds
flashing  On  1,6 mA  Sensor warm-up period  finishes automatically after 120 
seconds. 
10 
"
TEST
" 
flashing
Off
  glows  0 mA  Too low supply voltage  increase supply voltage 
11 
"
CHNLM
" 
"
SENS
"
flashing 
shortly 
(single 
flash) 
On  1,2 mA  Sensor lifetime expired  Replace sensor 
12 
"
CHEC
" 
" 
L  MAS
 "
Twin 
flash 
On  1,2 mA  After sensor replacement: 
Sensor is not specified for the gas 
Replace sensor again or 
select different gas; see page 19 
„Sensor Replacement“ 
13 
"
CHEC
" 
"
SCAL
"
Twin 
flash 
On  1,2 mA  After sensor replacement: 
Detection range setting of sensor and 
transmitter do not match for the gas 
Check and, if necessary, change 
detection range SCAL 
See page
 19 „Sensor Replacement
“
14 
" 
SYS
 "
"
ERR.1
"
Off  On  1,2 mA  Fault during RAM access 
Re-start transmitter. If error is 
reported again, replace transmitter
15 
" 
SYS
 " 
"
ERR.2
"
Off  On  1,2 mA  Fault during ROM access 
Re-start transmitter. If error is 
reported again, replace transmitter
16 
" 
SYS
 " 
"
ERR.3
"
Off  On  1,2 mA  Fault during EEPROM access (internal) 
Re-start transmitter. If error is 
reported again, replace transmitter
17 
" 
SYS
 " 
"
ERR.4
"
Off Off 1,2 mA Reserved 
Re-start transmitter. If error is 
reported again, replace transmitter
18 
" 
SYS
 " 
"
ERR.5
" 
" 
SYS
 " 
"
ERR.5
"
On  1,2 mA  Fault during regulation of sensor voltage
19 
Off
Off  On  1,2 mA  Fault during regulation of hardware 
zeropoint 
20 
" 
SYS
 " 
"
ERR.7
" 
Off On 1,2 mA Temperature measurement is not 
plausible 
Re-start transmitter. If error is 
reported again, replace transmitter
21 
"
SENS
" 
"
ERR.1
" 
"
SENS
" 
"
ERR.1
"
On  1,2 mA  Sensor is missing 
Fit sensor (automatic re-start) 
22 
"
SENS
" 
"
ERR.2
" 
Off  On  1,2 mA  Fault during EEPROM access (sensor) 
Replace sensor (automatic re-
start) 
23 
"
SENS
" 
"
ERR.3
" 
Off  On  1,2 mA  Wrong sensor/parameter memory 
24 
" 
ADv
 " 
"
ERR.1
" 
Off  On  1,2 mA  Error A/D converter (sensor voltage)  Re-start transmitter. 
If error is reported again, replace 
transmitter. 
25 
" 
ADv
 " 
"
ERR.2
" 
" 
ADv
 " 
"
ERR.2
"
On  1,2 mA  Error A/D converter (NTC)  Re-start transmitter. 
If error is reported again, replace 
transmitter. 
26 
" 
ADv
 " 
"
ERR.3
" 
Off
On  1,2 mA  Error A/D converter (sensor current)  Re-start transmitter. 
If error is reported again, replace 
transmitter. 