Connection Unit
Connection Unit Error Recovery
6
2-21 SL21 Cassette jam was 
detected in cassette 
load position #5. 
Cassette could be 
jammed by Stopper 
cylinder SL21. 
Connection 
Unit Top 
View
C-4 1. Press PAUSE/RUN to clear the error.
2. If the problem persists:
a. Open the right-side tabletop safety cover.
b. Ensure cassette is not trapped by Stopper 
cylinder SL21 at shuttle position #5.
3. If unable to remove trapped cassette:
a. Perform SHUTDOWN at Line Controller.
b. Power OFF automation line at Inlet Unit.
c. Cut off incoming lab air supply to the line. 
(If June-air compressor, turn Red Valve in 
back of Compressor Unit).
d. Manually depress air cylinder (SL21) and 
then remove trapped cassette.
e. Reposition the cassette in position #5.
f. Reconnect air supply.
g. Power ON the automation line at Inlet 
Unit.
h. Perform STARTUP at Line Controller.
4. If the problem persists, call your Beckman 
Coulter Representative.
2-22 SL22 Cassette may be 
jammed between 
cassette position #6 
and cassette position 
#7.
Connection 
Unit Top 
View
D-3 1. Open the left lower front cover.
2. Clear any obstructions if possible.
3. Remove jammed cassette and place into 
cassette position #7.
4. Replace covers.
5. Press PAUSE/RUN to clear the error.
6. If unable to remove trapped cassetteā
a. Perform SHUTDOWN at Line Controller.
b. Power OFF automation line at Inlet Unit.
c. Cut off incoming lab air supply to the line. 
(If June-air compressor, turn Red Valve in 
back of Compressor Unit.)
d. Manually depress air cylinder (SL22) and 
then remove trapped cassette.
e. Reposition the cassette in position #6.
f. Reconnect air supply.
g. Power ON the automation line at Inlet 
Unit.
h. Perform STARTUP at Line Controller.
7. If the problem persists, call your Beckman 
Coulter Representative.
Tab le 6. 1 Connection Unit Error Codes  (Continued)
Keypad 
Display Sensor
a
Problem
Sensor 
Diagram
Grid 
Ref Solution