Specifications
A64192AA2 FINAL
8-5
Stockyard Unit
Stockyard Unit Error Recovery
8
1-63 SN63 Proximity sensor did not 
detect a sample tube 
carrier on return lane.
Stockyard 
Unit T Lane
D--4  1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-64 SN64 Sample tube carrier 
error coming from H-
Lane incoming lane to 
make the turn to the 
return lane.
Stockyard 
Unit H Lane
B-4 1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-65 SN65 Proximity sensor did not 
detect a sample tube 
carrier turning on H-
Lane.
Stockyard 
Unit H Lane
B-4  1. Clear any obstruction.
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
1-67 SN67 Stockyard first level 
access door open.
3000 Tube 
Stockyard
To p  V i e w
C-4/5 1. Close the door.
2. Press PAUSE/RUN to clear the error.
3. If the error persists, call your Beckman 
Coulter Representative.
1-68 SN68 Stockyard second level 
access door open.
3000 Tube 
Stockyard
To p  V i e w
C-4/5 1. Close the door.
2. Press PAUSE/RUN to clear the error.
3. If the error purists, call your Beckman Coulter 
Representative.
1-69 SN69 Stockyard third level 
access door open.
3000 Tube 
Stockyard
To p  V i e w
C-4/5 1. Close the door.
2. Press PAUSE/RUN to clear the error.
3. If the error purists, call your Beckman Coulter 
Representative.
2-01 SL01 Stockyard Loading 
Shuttle Cup error on 
first level.
3000 Tube 
Stockyard
To p  V i e w
E-4/5
C
1. Ensure the Stockyard Loading Shuttle 
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-02 SL02 Stockyard Retrieving 
Shuttle Cup error on 
first level.
3000 Tube 
Stockyard
To p  V i e w
E-4/5
C
1. Ensure the Stockyard Retrieving Shuttle 
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-03 SL03 Stockyard Internal 
Tr a n s f e r  A r m  a i r  
cylinder error on first 
level moving up or 
down.
3000 Tube 
Stockyard
To p  V i e w
D-4/5
D
1. Ensure the Stockyard Internal Transfer Arm 
air cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
2-05 SL05 Stockyard Internal Arm 
Pushing cylinder error 
on the first level.
3000 Tube 
Stockyard
To p  V i e w
D-4/5
D
1. Ensure that Stockyard Internal Arm Pushing 
cylinder movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad 
Display Sensor
a
Problem
Sensor 
Diagram
Grid 