Specifications
A64192AA2 FINAL
8-8
Stockyard Unit
Stockyard Unit Error Recovery 
2-55 SL55 Stopper error at the 
T-Lane before sorting 
position.
Stockyard
Unit T Lane
C-3 1. Clear any obstruction. 
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-56 SL56 Stopper error at the 
T-Lane sorting position.
Stockyard
Unit T Lane
C-3 1. Clear any obstruction. 
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-57 SL57 Stockyard Divider Arm 
error at the entrance of 
H Lane.
Stockyard
Unit T Lane
C-3 1. Ensure the gate is free to extend and retract.
2. Ensure the gate arm is not to slow to extend/ 
retract. If so, the cylinder air flow controller 
may need adjustment.
3. Press PAUSE/RUN to clear the error.
2-58 SL58 Stopper error at the exit 
of the H-Lane.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction. 
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-59 SL59 Stopper error at the exit 
of the H-Lane.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction. 
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-60 SL60 Stopper error at the T-
Lane sample tube 
carrier sorting position 
#1.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction. 
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-61 SL61 Stopper error at the T-
Lane sample tube 
carrier sorting position 
#2.
Stockyard
Unit T Lane
C-4 1. Clear any obstruction. 
2. If a sample tube carrier is jammed, gently 
twist and push the sample tube carrier in the 
direction of least resistance.
3. Press PAUSE/RUN to clear the error.
2-62 SL62 Stockyard Divider Arm 
error at the T-Lane 
sample tube carrier 
sorting position #2.
Stockyard
Unit T Lane
C-4 1. Ensure the gate is free to extend and retract.
2. Ensure the gate arm is not to slow to extend/ 
retract. If so, the cylinder air flow controller 
may need adjustment.
3. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad 
Display Sensor
a
Problem
Sensor 
Diagram
Grid 
Ref Solution