Specifications
A64192AA2 FINAL
8-11
Stockyard Unit
Stockyard Unit Error Recovery
8
4-06 SM06 Failure to move the 
Stockyard Transfer Arm 
on the second level in 
the Y-axis direction.
3000 Tube 
Stockyard 
To p  V i e w  
C-4/5 1. Ensure the Stockyard Transfer Arm 
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Stockyard Unit, wait one 
minute and then Power ON the Stockyard 
Unit to reset.
b. Place Stockyard in AUTO mode to 
initialize.
4. If the problem persists, call your Beckman 
Coulter Representative.
4-07 SM07 Failure to move the 
Stockyard Transfer Arm 
on the third level in the 
X-axis direction.
3000 Tube 
Stockyard 
To p  V i e w  
D-4/5
D
1. Ensure the Stockyard Transfer Arm 
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Stockyard Unit, wait one 
minute and then Power ON the Stockyard 
Unit to reset.
b. Place Stockyard in AUTO mode to 
initialize.
4. If the problem persists, call your Beckman 
Coulter Representative.
4-08 SM08 Failure to move the 
Stockyard Transfer Arm 
on the third level in the 
Y-axis direction.
3000 Tube 
Stockyard 
To p  V i e w  
C-4/5 1. Ensure the Stockyard Transfer Arm 
movement is not obstructed.
2. Press PAUSE/RUN to clear the error.
3. If there is no obstruction:
a. Power OFF Stockyard Unit, wait one 
minute and then Power ON the Stockyard 
Unit to reset.
b. Place Stockyard in AUTO mode to 
initialize.
4. If the problem persists, call your Beckman 
Coulter Representative.
7-07 AS07 Stockyard Internal 
Transfer Arm gripper on 
the first level failed to 
pick up sample.
3000 Tube 
Stockyard 
To p  V i e w  
D-4/5
D
1. Ensure the sample tube is properly 
positioned in gripper fingers.
2. Press PAUSE/RUN to clear the error.
Tab le 8. 1 Stockyard Unit Error Recovery Codes (Continued)
Keypad 
Display Sensor
a
Problem
Sensor 
Diagram
Grid 
Ref Solution