Summary
Version 4.1.0 software provides product improvements for 
the Codonics Virtua
®
 Medical Disc Publisher.
This document details the changes from Version 4..0.0 
software. The Virtua User’s Manual, Reference Guide, 
Operating Software Release Notes, and Technical Briefs 
provide further information on the use of Virtua systems. If 
upgrading from an older software version, previous Virtua 
Release Notes can be downloaded from the Codonics website.
kÉï=cÉ~íìêÉë
This section details new features in Version 4.1.0 software.
◆
Anonymization.
 Virtua has been updated to provide 
Anonymization of patient data per DICOM Standard 
2011 3.15 E.2 Basic Application Level Confidentiality 
Profile. Each tag and method of anonymization may be 
custom configured as needed. Please refer to Technical 
Brief 901-459-001 found on your Accessories Disc or visit 
the Codonics web site for the latest publication of this 
and all other available user documentation.
◆
Viewer Updates.
 The following viewer updates are 
included in Virtua 4.1.0:
Clarity Viewer updated to v7.5 SP6
MIMvista Viewer updated to v1.2.4
Philips DICOM Viewer updated to v3.0
Removed Siemens Syngo Imaging XS Viewer
Added Siemens Plaza Viewer VA20D
Removed Invia Corridor Viewer from Accessories Disc
◆
JPEG Representation of DICOM Images. 
Virtua now 
provides JPEG representations of the DICOM images of a 
study. All original images are provided on the disc with 
the JPEG images generated. Please refer to Technical 
Brief 901-459-001 on your Accessories Disc or visit the 
Codonics web site for the latest publication of this and all 
other available user documentation.
Parameter: 
doCreateJpegImages
Settings: 
True | False
Default:  False
Description:  Controls whether Virtua produces JPEG 
version of the DICOM images in the 
study.
◆
DVD+R.
 Virtua now supports DVD+R discs for 
recording media.
◆
Virus Scan Support.
 Virtua now allows users to access 
all Virtua drives across a network. This permits a user to 
utilize a virus scanning software to inspect and alert the 
user of any suspected viruses on Virtua drives.
◆
Segmentation Storage DICOM Objects.
 Virtua now 
supports Segmentation Storage DICOM Objects.
◆
Added VNC (Virtual Network Computing) support to 
Virtua, which allows administrators to remotely 
connect to a unit within hospital networks or with VPN 
connections. 
TightVNC is deployed with the software 
installation and is set as an automatic startup Windows
®
service. The VNC full-control password is automatically 
set in synch with the unit’s administrative account 
password. 
The following parameters can be configured in the 
profiles\network\network.default.txt profile:
Parameter: 
vncFirewallPortOpen_
Settings: 
True | False
Default: 
True
Description:  When set to True, VNC is enabled.
Parameter: 
firewallEnabled
Settings: 
True | False
Default: 
False
Description:  When set to False, the firewall is disabled.
CAUTION: 
Version 4.1.0 software is intended for use on 
all Virtua systems except for the Virtua-1 model. Version 
4.0.0 software will not install on the Virtua-1 model 
(controller serial numbers beginning with 70C).
CAUTION: 
A full install is required when upgrading a 
standard Virtua system running 4.0.1 or previous 
software, during which all patient studies will be 
removed. Virtua XR/E/C will not require a full install to 
upgrade from previous software.
CAUTION: 
Newer Virtua systems, which utilize a 
separate print head and individual ink cartridges for each 
color, require Version 4.0.0 software or later to function 
properly. These systems can be identified by the first 
three characters in the serial number of the Recorder, 
83C
.
NOTE: 
VNC server settings configured on a unit will 
not be retained across software installations. Only the 
full-control password will be reset to synchronize with 