To initialize a temperature sensor to the Control Module, perform the
following procedure:
1. Disconnect all temperature sensors from the Control Module.
2. Disconnect the power from the Control Module. Wait for 10
seconds and then connect the power to the Control Module.
3. Connect the temperature sensor to the Control Module
4. Disconnect the power from the Control Module. Wait for 10
seconds and then connect the power to the Control Module.
5. If installing more than one temperature sensor, repeat steps 3
and 4 for each sensor added.
This procedure will initialize the temperature sensor(s) to the Control
Module. The Control Module identifies each temperature sensor
connected and initialized by an internal serial number. The Control
Module will define the sensors by the numbers 1 through 4 based on the
order in which they are initialized. Be sure to initialize the main outdoor
sensor first, so that it will be defined as T1. Wind chill, heat index and
dew point are all calculated based on the value of T1.