Only one alarm will be required for Door switches. The alarm for the door will be activated if
one of two conditions is met. 1) The door is open in a single continuous event for greater than
the set alarm delay. 2) If in a user-selected period of up to 1 hour, the alarm has been open
multiple times to a time greater value than the alarm time delay.
Acknowledgement of alarms via the Gateway login, including corrective action via text box.
(21 CFR Component)
Audit trail of events. (21 CFR Component)
Electronic Signatures of Data before printing (21 CFR component) TBC
Ability to see current data from last 24 Hours or 7 days and 28 days in graphical format.
Graphing of multiple channels on one graph by transmitter.
Ability to load and view historical data.
The connection of any screen/keyboard and mouse will be disabled
Ability to e-mail selected contacts with alarm or channel information either summary or data.
Low battery warning from individual transmitter or repeater
The graphical display of Alarm events on the Web Browser summary page or individual
channel screen will highlight the current status of alarms and alarm acknowledgements. The
name of the person acknowledging the alarm and the corrective action can also be recorded.
Indicator of nodes out of radio contact, indicated as ‘No Signal’.
Summary for each channel/node showing current status of the node. This includes MAX, MIN
and Average readings for active channel, excluding any under or over range values. Current
status of alarms for each channel and indication of historical alarm events for each active
channel. MKT and Lethality calculations can be programmed in the Task for each external
temperature channel.
GATEWAY Screen Options
The Gateway screen options are split by category.
Administration
Administration includes the functions listed below…..
Users – Setup, reinstatement or removal of current/new users/administrators
Gateway Name – Description of Gateway. (Important if more than one in use.)
Locations list – Add up to 32 locations for the transmitters.
Setup Switch Output – Alarm output relays set to either NO or NC
Set Clock – Set date and time